Philips Respironics BiPAP Vision Features

The Respironics BiPap Vision Ventilator is specifically designed to address the unique requirements of Invasive Ventilation within the hospital, including changing leaks, dynamic breathing patterns, and high FIO2 and flow demand. The BiPAP Vision features accurate monitoring, reliable alarms, and an intuitive menu. Auto-Trak ensures optimum sensitivity despite changing breathing patterns or circuit leaks. Auto-adaptive triggering and cycling reduce work of breathing and support patient-ventilator synchrony, supporting patient compliance and reducing the need for manual adjustments. The Vision is designed with advanced technology to be accurate, reliable, and easily upgradeable. This low-cost alternative form of sub-acute therapy may reduce the need for intubation for certain patients.

  • Digital Auto-Trak sensitivity automatically adapts to changing leaks on a breath-by-breath basis, allowing the patient’s mask to be fit for comfort rather than a perfect seal.
  • The integrated display screen displays real-time graphics in waveform or bar scale format to assist with ventilator management.
  • CPAP and S/T modes are available for maintaining airway patency and providing pressure support ventilation.
  • Oxygen control allows 21–100% concentrations from without attaching auxiliary oxygen to the mask
  • Pressure regulation at the mask ensures accurate therapy for the patient.
  • The alarm module provides a complete range of alarms that are designed with mask ventilation in mind, minimizing nuisance alarms caused by mask leak.

Integrated Display Screen displays easy-to-read real-time graphics in waveform or bar scale format, providing the following measured and calculated numerical values to assist with ventilator management:

  • IPAP
  • Total Breath Rate
  • Exhaled Minute Ventilation
  • % Patient Triggered Breaths
  • Exhaled Tidal Volume
  • EPAP
  • Ti/Tot
  • Peak Inspiratory Pressure
  • Patient or Total Leak
  • Integrated Display Screen
  • Oxygen Control up to 100%
  • Pressure Regulation at Mask
  • BiPAP
  • Auto-Trak Sensitivity
  • Alarm Module


Description

Philips Respironics BiPAP Vision Specifications

Dimensions

  • At the Base
  • Height: 10.625 in (27 cm)
  • Width: 10.375 in (36.5 cm)
  • Length: 16 in (40.6 cm)
  • Weight: 34 lbs (15.4 kg)

Electrical

  • AC Input Voltage (VAC):
    • 100/120/230/ 240 VAC
    • Single Phase
    • ±10%
  • Fuses:
    • 100 – 120 VAC ~ T 3.5 A, 5 × 20 mm, Time Lag (×2) (For serial no.’s 100500 and higher – Respironics Reorder # 1000749)
    • 115 VAC ~ T 3.0 A, 250 V, ¼” × 1¼” (For serial no.’s 100499 and lower – Respironics Recorder # 582100)
  • 220 VAC, 230 VAC and 240 VAC ~ T 1.6 A, 250 V, 5 × 20 mm (For all serial no.’s – Respironics Recorder # 1000750)
  • Power Consumption: 300 VA max.
  • AC Current: 3.0 A maximum
  • AC Frequency: 50/60 Hz

Modes

CPAP

  • Control Range: 4 to 20 cm H2O
  • Control Increments: 1 cm H2O

S/T

  • Control Range: 4 to 40 cm H2O
  • Control Increments: 1 cm H2O

PAV/T

  • Control Range: Quick Start Menu
  • Control Increments: Obstructive, Restrictive, Mixed, Normal
Additional information
MARCA

Philips Respironics

CONDICIÓN

Reacondicionado Certificado

Más Información

Más Información

Control Ranges & Increments: Parameters

IPAP

  • Control Range: 4 to 40 cm H2O
  • Control Increments: 1 cm H2O

EPAP

  • Control Range: 4 to 40 cm H2O
  • Control Increments: 1 cm H2O

CPAP

  • Control Range: 4 to 40 cm H2O
  • Control Increments: 1 cm H2O

Rate

  • Control Range: 4 to 40 BPM
  • Control Increments: 1 BPM

Timed Inspiration

  • Control Range: 0.5 to 3.0 sec
  • Control Increments: 0.1 sec

IPAP Rise Time

  • Control Range: 0.05 to 0.40 sec
  • Control Increments: 4 set points: .05, .1, .2, .4 sec

Oxygen Concentration (%O2)

  • Control Range: 21 to 100%
  • Control Increments: 4% from 21 to 25% 5% from 25 to 100%

Control Ranges & Increments: Alarms

High Pressure

  • Control Range: 5 to 50 cm H2O
  • Control Increments: 1 cm H2O

Low Pressure

  • Control Range: Disabled; 0 to 40 cm H2O
  • Control Increments: 1 cm H2O

Low Pressure

  • Control Range: Delay 0 to 60 sec
  • Control Increments: 1 sec

Apnea

  • Control Range: Disabled; 20 to 40 sec
  • Control Increments: 4 set points: Disabled, 20, 30, 40 sec

Low Minute Ventilation

  • Control Range: Disabled; 1 to 99 L/min
  • Control Increments: 1 L/min

High Rate

  • Control Range: 4 to 120 BPM
  • Control Increments: 4 set points: 1 BPM

Low Rate

  • Control Range: 4 to 120 BPM
  • Control Increments: 4 set points: 1 BPM

Display Ranges & Increments

IPAP

  • Display Range: 0 to 50 cm H2O
  • Display Resolution: 1 cm H2O

EPAP

  • Display Range: 0 to 50 cm H2O
  • Display Resolution: 1 cm H2O

CPAP

  • Display Range: 0 to 50 cm H2O
  • Display Resolution: 1 cm H2O

Rate

  • Display Range: 0 to 150 BPM
  • Display Resolution: 1 BPM

Exhaled Tidal Volume (VT)

  • Display Range: 0 to 4000 ml
  • Display Resolution: 1 ml

Minute Ventilation (Min Vent)

  • Display Range: 0 to 99 L/min
  • Display Resolution: 1 L/min

Total Leak (Tot Leak)

  • Display Range: 0 to 300 L/min
  • Display Resolution: 1 L/min

Patient Lead (Pt. Leak)

  • Display Range: 0 to 300 L/min
  • Display Resolution: 1 L/min

Peak Inspiratory Pressure (PIP)

  • Display Range: 0 to 50 cm H2O
  • Display Resolution: 1 cm H2O

Percent of Pateint Triggered Breaths (Pt. Trig)

  • Display Range: 0 to 100%
  • Display Resolution: 1%

Ti/Ttot

  • Display Range: 0 to 100%
  • Display Resolution: 1%